- ページ 9

ドライヤー DRY AGER DX 500のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。DRY AGER DX 500 17 ページ。
DRY AGER DX 500 にも: ユーザーマニュアル (14 ページ), 交換手順 (2 ページ)

DRY AGER DX 500 取扱説明書
Control ON and OFF
Press the button once for 5 seconds and the OFF
symbol is displayed. The LED 4 (= standby)
The compressor is turned off. All relays are deactivated.
Press the button again and the compressor will restart.
Locking and unlocking Keypad
1. Press
and
of 3 seconds, POF appears for a few seconds
(POF = OFF parameter)
2. Thereafter, only displayed setting can not be
changed. Furthermore, the MIN and MAX values
can be viewed.
further activated.
UNLOCK
Press
and
hold the two keys for a minimum of 3 seconds.
Setting temperature
1. Press SET key: The set value is displayed and the
LED's flash
set
temp.
2. Now to change the settings, within 10 seconds,
use the
and
3. To confirm press SET button or wait 10 seconds
Start manual defrost
Hold down button for a min. 2 seconds
The manual defrosting starts.
Is indicated through LED3.
Turn On / Off UVC Desinfection
Press the button for at least 5 sec.
Is indicated through LED2.
and hold together, for a minimum
The ON / OFF switch can be
arrow keys

Error Messages

Reasons for
Message
error messages
"P1"
Room sensor error
Evaporation
"P2"
temperature
sensor Failure
"P3"
Humidity Sensor error
"HA"
High temperature alarm
"LA"
Low temperature alarm
"HHA"
HIGH humidity alarm
"LHA"
LOW humidity alarm
"dA"
Door alarm
Displays the alarm message until the alarm conditions exist. All
alarms are displayed alternately with the room temperature except
for the P1. EE can be acknowledged by pressing any key "rSt"
for 3 seconds in the display, then returns to normal operation.
Acknowledgement of the acoustic alarm
The audible alarm can be (optional) switched off by
pressing any key.
Alarm acknowledgement
Message P1, P2 and P3 after about 30 seconds; after correction
or troubleshooting, the error message disappears automatically.
Check probe sensor before replacing. Check connections.
Temperature alarms HA and LA will automatically reset once the
normal temperature is reached or when the defrost starts. DA
goes off when the door is closed. External alarms EAL and BAL
are disabled after the digital input. PAL acoustic alarm: Through
Device Turning off or standby ON / OFF.
Humidity alarms HHA and LHA are acknowledged automatically
when normal range has been reached.
Maintenace information
Information
Reason
Lifetime of UVC-bulb and active
"UVC"
carbon filter catridge are reached.
Please order the Dry Ager Service Kit and change the UVC bulb,
Art. DX0010 and Active Carbon Filter Catridge, Art. DX00110.
After exchaning you can reset the information:
and SET % RH and hold the keys for a minimum
1. Press
of 2 sec.
2. Select with button
Parameter UcH.
3. Press button SET % RH. The set value is displayed and
value is flashing.
4. With button
set the value to „0".
5. Confirm with button SET % RH.
6. After 1 minute the display switches automatically in
standard display.
Outputs
Cooling and Heating OFF
Defrost by time
moisture control OFF
regulation unaffected
regulation unaffected
regulation unaffected
regulation unaffected
regulation unaffected
9