SUGGESTED PROCEDURE FOR OBTAINING
STERILE WATER SAMPLES:
Prior to taking the water sample, be sure to have on hand an
adequate supply of sterile bottles (as shown in Fig 1). These
sterile bottles should be obtained from a reputable laboratory
and should have been autoclaved and contained within a
plastic outer wrapping.
1. Prior to taking the sample, it is imperative that the sample
valve, faucets, etc. be opened at full force for a complete three
and a half minutes.
2. After the valve has been fully opened for three and one
half minutes, reduce the flow to a reasonable stream of
water. Flow to drain an additional three minutes.
3. Open the sterile bottle or sterile container being used.
Holding the cap in a down position, the operator should then
hold his breath while taking the sample so as to avoid oral
contamination of the sample. The operator must not allow his
finger to touch the inside of the cap or the neck of the bottle.
4. After the sample has been taken, the cap should be tightly
placed on the sample container.
5. The sample container should be placed in a plastic wrapping,
kept cool, and should be taken to the laboratory for testing as
soon as possible following the above procedure.
We recommend duplicate samples be taken at each test station
during each specific test so as to avoid loss of sample through
laboratory error and to insure reasonable validity through
comparison.
Check with your local laboratory to assure proper sampling
and submittal procedure.