Electromagnetic emissions:
B‐ONE is intended for use in the electromagnetic environment specified below. The B‐ONE should only be used in such an environment.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000‐3‐2
Voltage fluctuation/ flicker
emissions
IEC 61000‐3‐3
Electromagnetic Immunity:
B‐ONE is intended for use in the electromagnetic environment specified below. The B‐ONE should only be used in such an
environment.
Immunity test
6kV contact
Electrostatic discharge
8kV air
(ESD) EN61000‐4‐2
2kV for power supply
Burst/Fast transient
EN 61000‐4‐4
lines
1kV differential mode
Surge
2kV common mode
EN 61000‐4‐5
Immunity test
5%Ut
Voltage dips, short
interruptions and
(95% dip in U
voltage variation on
cycle
power supply input
lines.
40% Ut
(60% dip in U
70% U
(30% dip in U
cycle
5% U
EN 61100‐4‐11
(95% dip in U
Power frequency
(50/60 Hz)
magnetic field
3A/m
EN 61100‐4‐8
B‐ONE is intended for use in the electromagnetic environment specified below. The B‐ONE should only be used in such an
environment.
USER MANUAL Rev 25/05/2023
15. ACCOMPANYING DOCUMENTS
Group 1
B‐ONE uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class B
Class A
B‐ONE is suitable for use in all establishment including domestic establishments and
Conforms
those directly connected to the public low‐voltage power supply.
Conforms
Test level
EN 60601‐1‐2
6kV
8kV
2kV for power supply
lines
1kV differential mode
2kV common mode
Test level
ICE 60601
5%Ut
) for 0,5
(95% dip in U
t
cycle
40% Ut
) for 5 cycle
(60% dip in U
v
cycle
t
) per 25
70% U
t
(30% dip in U
cycle
t
5% U
) for 5 sec
t
3A/m
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]
B‐ONE
Page 13 of 16
Compliance level
Floor should be wood, concrete, or ceramic tile. If
floor is covered with synthetic material, the relative
humidity should be at least 30%
Quality of the Main power source should be that of a
typical commercial or hospital environment
Quality of the Main power source should be that of a
typical commercial or hospital environment
Compliance level
Quality of the Main power source should be that of a
typical commercial or hospital environment If the
) for 0,5
t
user of the B‐ONE requires continued operation
during power main interruptions, it is recommended
than the B‐ONE be powered from an uninterruptible
power supply or battery.
) for 5
v
t
) per 25
t
t
(95% dip in U
) for 5
t
sec
Power frequency magnetic fields should be at a
typical commercial or hospital establishment level.
Electromagnetic environment ‐ guidance
Electromagnetic environment ‐ guidance