5.

Calibration

Options
Display
Enter
Calibration
Mode
SETUP
SCALE
Full Scale
_XXXX
Resolution
E X
Decimal
Pt X
Point
Zero
Calibration
LoAd0
-----
Load1
Calibration
_XXXX
-----
Load2
Calibration
_FULL
-----
_End_
10 |
ANYLOAD OCSL Mini Type Crane Scale User's Guide
Operations
Hold and press [HOLD] and
[TARE] simultaneously, keep
pressing until SETUP
displays.
Press [ON/OFF], [HOLD] and
[TARE] simultaneously, keep
pressing until SCALE
displays.
Press [TARE].
Press [HOLD] to change
Full Scale and press [TARE]
to save setting.
Press [HOLD] to change
Resolution and press
[TARE] to save setting.
Press [HOLD] to change
Decimal Point and press
[TARE] to save setting.
Press [TARE] to start zero
calibration.
Press [HOLD] to change 1st
Calibration Weight value.
Load the actual weight then
press [TARE] to start span
calibration.
nd
Load the 2
test weight
then press [TARE] to finish
the calibration.
Notes
Enter Setup Mode
(At this stage set the unit of
measurement you are going to
calibrate, hold & press
the[HOLD/UNIT] to switch unit)
Enter Calibration
Mode
For example, if you set
the unit at kg and setting
the parameter here to
600 then the full scale
should be 600kg else if it
was set to lb then the full
scale shall be equivalent
to 600lb.
1/2/5
0: XXXXX
1: XXXX.X
2: XXX.XX
3: XX.XXX
Set the scale to no load
status and must be at stable
state.
If using test weight
equivalent to full scale
(FULL), the calibration shall
be finished at this stage.
nd
If 2
test weight is not
needed, you can skip this
stage by pressing the
[ON/OFF] to finish and exit
the Calibration.