Apex2
Calibrate the pump
Always calibrate the pump at the sampling flow rate for greatest accuracy.
Note: If using a filter with a small pore size or if operating at a high flow rate,
which results in a high back pressure, it may take the pump a little longer to
stabilise when adjusting the calibrated value.
1. Attach a sampling head or flow tube or other flow-measuring device to
the pump's inlet nozzle.
2. Navigate to CALIBRATE.
3. Press
4. Press
speed until the meter flow rate and the pump flow rate match.
5. Press
Start/stop a sampling run
Ensure your pump is calibrated, set to the correct flow rate and you have reset
the accumulated data if necessary.
To start a sampling run:
1. Turn on the pump.
2. Press
The following screens should be visible.
To stop or pause a sampling run:
Press
to start the test.
or
to change the pump
to stop the test.
for 3 seconds.
for 3 seconds.
3