- ページ 8

試験装置 Additel 927のPDF 整備・修理マニュアルをオンラインで閲覧またはダウンロードできます。Additel 927 15 ページ。

Additel 927 整備・修理マニュアル
⑤ Please
refer to picture 2.1.3 to remove the supporting foot, end cover, plastic liquid reservoir and parts 12 and 13. Use a 22 mm wrench
to remove the pre-pressurization cylinder and part 14.
9.End cover
8.Supporting foot
Exchange the damaged parts, and reassemble the parts in reverse order.
Pre-pressurization piston parts list.
05
10.Plastic liquid reservoir
12
13
11.Pre-pressurization
Cylinder
Picture.2.1.3
12
14
Need to add
thread-locking adhesives