Step 5:
Step Rate
Ask patient to sit to allow the vacuum level to
drop to the low activity threshold. Once vacuum
level has settled to low threshold, watch the
Dashboard and ask the patient to walk at a slow
pace. The vacuum level should stay in the low
threshold vacuum level. Next, ask the patient to
walk a medium pace. The vacuum level should
jump to the medium activity level threshold.
Next, ask the patient to walk at a fast pace. The
vacuum level should jump to the high activity
threshold. See images below.
Slow Pace
Walking
Medium Pace
Walking
Fast Pace
Walking