Setup
iQ Vacuum is designed to not require additional build height by
clamping on to the prosthesis pylon or adapter. The amount of height
required between the socket and the foot depends upon the top profile
of the prosthetic foot chosen.
Pylon Attachment
Remove the front cover screw to reveal the clamp. Remove the two
clamp bolts. Select a rubber clamp bushing from the included kit to
wrap around the pylon or adapter. The clamp on the main body of
the iQ Vacuum unit should fit snuggly against the pylon and rubber
shim. Position the unit as high as possible on the posterior side of the
prosthesis pylon. Take care that there is no contact with the socket or
the foot when mounted. If desired, the provided self-adhesive foam
pylon bumper should be attached at the bottom of the housing below
the serial number label. Attach the anterior clamp with the two clamp
bolts and tighten with a torque wrench to 5 Nm. Check that the unit
is held firmly and does not slip. You may have to choose a different
thickness of rubber clamp bumper if it is not secure, or if you cannot
tighten the clamp fully. If needed, you can attach the unit to any side
of prosthesis for a better fit.
8
Freedom Innovations |
iQ Vacuum
| Instructions for Use
Maintenance & Cleaning
iQ Vacuum was designed for no internal service requirements for
the life of the device.
In order to maintain the IP67 rating, the external case of the iQ Vacuum
should never be open except for authorized service. The Water Mode
should always be activated when entering the water.
The external filter of the iQ Vacuum should be replaced every 6 months.
The iQ Vacuum Pump may be cleaned by wiping it using a soft cloth
with mild soap and water. Do not use any other solvents. Do not
immerse to clean.
Vacuum Connection and Filter
With iQ Vacuum securely fastened to the prosthesis, attach the filter
assembly with 3,2 mm tubing to the socket barb fitting, which should
be as close to the bottom of the socket as practical.
Filter Assembly
Cut Vacuum Line
Connect the 1,6 mm tubing to the filter assembly and route it to the
proximal barb on the iQ Vacuum pump. Ensure the tubing passes
through the proximal relief cutout in the housing. Cut the tubing to
length as necessary.
9