医療機器 Freedom Innovations PLIE 3のPDF プログラミング・リファレンス・マニュアルをオンラインで閲覧またはダウンロードできます。Freedom Innovations PLIE 3 2 ページ。
Programming
Reference Guide
Program Plié 3 in
5 Minutes or Less
in 5 Easy Steps!
The Plié Control 6 software is required to program a Plié 3
MPC Knee. To download the software, please visit the Freedom
Innovations website at: www.pliesupport.com/download
1
Connect
•
Remove and reinsert the battery
to turn on the internal Bluetooth
radio in knee.
•
Click Search. Once the serial
number corresponding to the
Plié is visible,
click Connect.
Calibrate
•
In the Setup Wizard window on
the right,
instructions in the window to check
if calibration is required, and perform
calibration if needed.
2
click Calibrate.
Follow the
10 mm
10mm
Refer to Troubleshooting Tips in Table 3
if you see the knee drift into flexion while
standing or a knee bobble at midstance.
Load Optimal
Settings
Enter or load optimal settings based on your patient's
activity level found in Table 1. These settings establish thresh-
olds for swing release and stumble recovery parameters for
proper knee function for your patient.
the Plié Control middle window. Type values into the box for
appropriate setting. Or, load previously saved files by
File > Load Knee Settings
and selecting the appropriate file.
Table 1: Plié Control Program Settings
Setting
Parameters
Low K3
Toe Threshold (TT)
400
Swing Transition (ST)
400
Heel Threshold
3000
Algorithm Delay Time
105
Flexion Angle Timer
400
Pre-Swing Timer
100
Maximum Heel Rise
55
Maximum Swing Angle
1000
Maximum Open Spool
600
Minimum Open Spool
30
Torque Boundary Limit
85
Maximum Swing Torque
270
Align
Plumb Line from bisected
socket should fall 0-5 mm
anterior to knee center
3
Click
More Params
clicking
Recommended Value
Mod K3
High K3
500
800
500
800
3000
3000
105
105
250
200
100
150
55
55
2200
2600
500
500
30
30
130
150
300
350
in