OPERATIONS

Polarimeter should be placed on a stable, vibration free surface in a dry & clean room with
good ventilation & temperature around 20 .
Power supply to the instrument is 230V and should be sourced through voltage stabilizer 1KVA
with a proper earthing.
Turn on the power supply switch (Allow sufficient warm up time ‐ 20 min. after and before
testing to use the instruments.)

KEYBOARD DESCRIPTION

0~9 : Numeric keys for entry of all parameters
AUTO : For Automatic measurement of parameters (measurement for the number of times
automatically as per parameter N.
MANUAL : For manual measurement of the parameter
SET UP / ◄ : Setting the parameters, Stopping the test, Back entry of parameters.
ENTER : Forwarding of parameters in the entry mode, starting of test.
CLEAR : Clearing the screen.
ZERO : Zeroing the reading.

DISPLAY DESCRIPTION

MODE : There are 4 modes of measurement corresponding to different measuring units.
MODE 1 : Optical Rotation
MODE 2 : Specific Rotation
MODE 3 : Concentration
MODE 4: International Sugar Scale
L is the length of sample tube in decimeter. There are 2 sizes of tubes available L = 1 dm(10 cm)
& L = 2 dm(20 cm)
C is concentration of solution in gms/100ml.
N is number of measurements.
SET : For setting parameters by pressing ENTER key with display SET highlighted.
OK : OK of parameters when ENTER Key is pressed with OK highlighted