- ページ 6
アクセサリー Apogee S2-131-SSのPDF オーナーズマニュアルをオンラインで閲覧またはダウンロードできます。Apogee S2-131-SS 15 ページ。 Red - far-red sensors
SPECIFICATIONS
Power Supply
*Output (sensitivity)
*Calibration Factor
(reciprocal of sensitivity)
Calibration Uncertainty
*Output Range
Measurement Repeatability
Long-term Drift
Non-linearity
Response Time
Field of View
Wavelength Ranges
Directional (Cosine)
Response
Temperature Response
Housing
IP Rating
Operating Environment
Dimensions
Mass
Cable
*The Ouput, Calibration Factor, and Output Range are variable from sensor to sensor, and the specifications listed are typical
values.
Calibration Traceability
Apogee S2 series Red - Far-Red sensors are calibrated through side-by-side comparison to the mean of three
transfer standard sensors under a quartz halogen lamp. The transfer standard sensors are calibrated through side-
by-side comparison to red and far-red photon flux density calculated from solar spectra collected in Logan, Utah,
using an Apogee model PS-300 spectroradiometer. The PS-300 spectroradiometer is calibrated with a quartz
halogen lamp traceable to the National Institute of Standards and Technology (NIST).
0.08 mV per µmol m⁻² s⁻¹ (typical value, variable from sensor to sensor)
12 µmol m⁻² s⁻¹ per mV (typical value, variable from sensor to sensor)
Anodized aluminum body with acrylic diffuser
5 m of shielded, twisted-pair wire; TPR jacket (high water resistance, high UV stability, flexibility
in cold conditions); pigtail lead wires; stainless steel (316), M8 connector
S2-131
Self-powered
± 5 %
0 to 33 mV
Less than 1 %
Less than 2 % per year
Less than 1 % (up to 400 µmol m⁻² s⁻¹)
Less than 1 ms
180°
645 to 665 nm ± 5 nm (Red)
720 to 740 nm ± 5 nm (Far-red)
± 2 % at 45°; ± 5 % at 75° zenith angle
Less than 0.1 % per C
IP68
-40 to 70 C; 0 to 100 % relative humidity
30.5 mm diameter, 36.6 mm height
51 g