OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying or
recording a millivolt (mV) signal (an input measurement range of approximately 0-250 mV is required to cover the entire
range of total shortwave radiation from the sun). In order to maximize measurement resolution and signal-to-noise ratio, the
input range of the measurement device should closely match the output range of the pyranometer.
SP-110: The sensor is self-powered and applying voltage will damage the sensor.
VERY IMPORTANT:
Apogee changed all wiring colors of our bare-lead sensors in March 2018 in conjunction
with the release of inline cable connectors on some sensors. To ensure proper connection to your data device,
please note your serial number or if your sensor has a stainless-steel connector 30 cm from the sensor head then
use the appropriate wiring configuration below.
Wiring for SP-110 Serial Numbers range 0-60050
Wiring for SP-110 Serial Numbers 60051 and above or with a cable connector
Red: Positive (signal from sensor)
Black: Negative (signal from sensor)
Clear: Shield/Ground
Black: Negative (signal from sensor)
Clear: Shield/Ground
White: Positive (signal from sensor)