- ページ 11

アクセサリー Apogee Instruments SU-202のPDF オーナーズマニュアルをオンラインで閲覧またはダウンロードできます。Apogee Instruments SU-202 18 ページ。 Ultraviolet-a sensor
Apogee Instruments SU-202 にも: オーナーズマニュアル (18 ページ)

Apogee Instruments SU-202 オーナーズマニュアル

OPERATION AND MEASUREMENT

Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying
or recording a millivolt signal. An input measurement range of approximately 0-10 mV is required to cover the
entire range of the SU-200 UV-A sensor. In order to maximize measurement resolution and signal-to-noise ratio,
the input range of the measurement device should closely match the output range of the UV sensor. DO NOT
connect the sensor to a power source. The sensor is self-powered and applying voltage will damage the sensor.
Wiring for SU-200
White: Output Signal
Black: Ground
Clear: Shield