- ページ 12

計測機器 Apogee Instruments SIF-111のPDF オーナーズマニュアルをオンラインで閲覧またはダウンロードできます。Apogee Instruments SIF-111 17 ページ。 Infrared radiometer
Apogee Instruments SIF-111 にも: オーナーズマニュアル (19 ページ)

Apogee Instruments SIF-111 オーナーズマニュアル
Sensor Calibration
Apogee SIF series infrared radiometers are calibrated in a temperature controlled chamber that houses a custom-
built blackbody cone (target) for the radiation source. During calibration, infrared radiometers (detectors) are held
in a fixture at the opening of the blackbody cone, but are thermally insulated from the cone. Detector and target
temperature are controlled independently. At each calibration set point, detectors are held at a constant
temperature while the blackbody cone is controlled at temperatures below (12 C), above (18 C), and equal to the
detector temperature. The range of detector temperatures is -15 C to 45 C (set points at 10 C increments). Data
are collected at each detector temperature set point, after detectors and target reach constant temperatures.
All Apogee analog infrared radiometer models (SIF-100 series) have sensor-specific calibration coefficients
determined during the custom calibration process. Unique coefficients for each sensor are provided on a
coefficient certificate (example shown below).
Calibration overview data are listed in the box in the upper left-hand corner, sensor specific coefficients are
listed in the box in the upper right-hand corner, temperature errors are shown in graph, and calibration date is
listed below descriptions of calibration procedure and traceability.