Fault Finding
1. FAULT
Trolley does not respond to input from the handset. No audible beeps from control box when buttons are pressed.
Fix
Step 1
Check that the EMERGENCY STOP button is not 'on'.
Step 2
Check that the battery is charged.
Step 3
Check that all the cables are plugged in correctly.
2. FAULT
Trolley does not respond to input from the handset. Audible beeps from control box when buttons are pressed.
NOTE: Battery must be fully charged before attempting the following procedures.
Fix
Step 1
Perform an ERROR RESET. Simultaneously press and hold the number 2 button until the intermittent
beeping stops. If this does not fix the problem contact your service provider or 4Healthcare for directions.