7.

Care and maintenance

7.1.
If the unit requires maintenance, disconnect the unit from the mains and contact your
local representative.
7.2.
All maintenance and repair operations must be performed only by qualified and
specially trained personnel.
7.3.
Fuse replacement.
7.3.1. Fuse for the control box. Disconnect the unit from electric circuit. Remove power cable
from its socket on the rear side of the unit. Open the fuse holder by pulling out the
holder (fig. 4, compact models) or by screwing off the holder lid marked FU2 (fig. 6,
model UVT-S-AR). Check the fuse and replace if necessary, М 3.15 A (type M - time
lag: Medium).
7.3.2. Fuse for the mains outlets (for models with built-in mains outlets). Disconnect the unit
from electric circuit. Open the fuse holder by screwing off the holder lid (fig. 5/2 or fig
6, FU1). Check the fuse and replace if necessary, М 5.0 A (type M - time lag: Medium).
Figure 4.
7.4. UV lamp replacement. Average lifetime of UV lamps supplied is 9000 hrs. Replacement
is necessary after lamp stops functioning or at the end of manufacturer specified
lifetime.
Note. Open UV lamp operation can be checked with the visible light lamp switched
off (switch 1 is OFF). Use the indicator in the centre of recirculator cover (fig.
1/3) to check operation of UV lamp inside the recirculator. If indicator lights
from inside while switch 2 is ON, then the UV lamp is functioning.
7.5. Dust filter replacement. The dust filters (fig. 3/2) on either end of the UV recirculator
with the hidden UV lamp should be checked monthly and cleaned or replaced when
they become clogged. To check, replace or clean the filters, simply unclip the covers,
fit a new filter if necessary or rinse current filters in water, dry and place back. Clip
covers back in place.
Figure 5
Figure 6
13