CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
23.
F
Ensure all breakers (main power supply and sub-breakers) are still in the OFF position. Return
power to the chamber by turning ON the main power supply breaker from building electrical
service. Ensure that ON/OFF switch is in the OFF position if being used. Turn chamber power
supply sub-breaker ON and observe the controller display as it boots. If the controller does not
boot to the Select Chamber screen, check the wiring and terminals at the rear of the display.
Ensure wiring is correct and the terminal connectors are properly inserted into their corresponding
receptacles.
24.
F
If the display still does not boot properly after re-checking power and communication connections,
call Conviron Technical Support (see Page iii).
25.
F
Switch all sub-breakers in the control panel to the ON position and begin the commissioning
procedures required to verify chamber operations.
26.
F
The configuration file uploaded to the controller at the factory is specific to the chamber in
which the CMP6050 retrofit control panel is being installed. To confirm the correct file has been
uploaded, tap the information icon
serial number for which the controller configuration file was programmed (if the information icon is
not visible, tap the exit icon
serial number at the controller matches the serial number on the exterior of the chamber. If it does
not, contact Conviron Technical Support (see Page iii).
27.
F
Input information retrieved in Step 1 to the new CMP6050 controller. The chamber is now ready to
be used.
10
May 2013 |
257775R00
in the upper left corner of the main status screen to view the
as necessary to return to the main status screen). Ensure that the
(END)
CONVIRON