- ページ 6
エアーコンプレッサー AUSTIN SCIENTIFIC M600のPDF マニュアルをオンラインで閲覧またはダウンロードできます。AUSTIN SCIENTIFIC M600 13 ページ。
RETURN HELIUM PRESSURE
SUPPLY HELIUM PRESSURE
COOLANT DATA
WEIGHT
3.2 OPERATING AND COOLANT CONDITIONS
When the compressor is operating, the ambient temperature must
remain between 40°F and 104°F (5°C and 40°C).
The standard connections on this compressor, for the coolant system,
are 3/8" Swagelok connectors. Connections can be modified to meet
customer requirements.
Heat exchanger manufacturer recommends 1200-840 micron strainer
on water supply and the following water requirements:
FLOW RATE (MIN) GPM
MIN. INLET TEMPERATURE °F(C)
MAX. INLET TEMPERATURE °F(C)
3.2.1 OPERATING CONDITIONS:
3.2.2 COOLANT CONNECTIONS:
3.2.3 COOLANT QUALITY
Electrical conductivity: >50 µS/cm
Iron:
Manganese:
Ammonia:
pH value:
Chloride:
Free carbon acid:
Nitrate:
Sulphate:
Sulphide:
50HZ PSI (bar)
60HZ PSI (bar)
50HZ PSI (bar)
60HZ PSI (bar)
Lb(kg)
<1.5 mg/l
<0.5 mg/l
<2 mg/l
6 – 9
<100 mg/l
<20 mg/l
<100 mg/l
<50 mg/l
0 mg/l
50
50
50
50
280
280
300
300
1.5
1.5
40(5)
40(5)
85(30)
85(30)
230(105)
230(105)