Other Symptoms: Fails To Cure Adhesive In Allotted Time
Possible Cause
Lamp Beyond Useful Life
Transmission loss In Light
Guide Too Great
Transmission loss In Filter
(Blue Filter At End Of Light
Guide) Too Great
Contaminants On Light
Guide Or Light Guide Filter
Lamp/Reflector Assembly
Not Installed Properly
TM-021 PN 35933
Rev 05/06/97
Problem: Low Output Intensity
Testing
Use a radiometer (model Dymax
Accu-Cal 20) to measure actual
output intensity. Consult manual
for proper output.
Compare light guide output against
new light guide (or use the Dymax
Light Guide Simulator - P/N
36987) to determine transmission
loss.
Compare filter output against new
filter to determine transmission
loss.
Visually examine ends of light
guide and filter for contaminants.
Visually check to make sure the
lamp/reflector assembly is seated
flush in the lamp mount assembly
(any error in installation could
cause a low output).
TROUBLESHOOTING
Corrective Action
Replace lamp/reflector
assembly if beyond useful
life (typical = 1000 hours).
Replace light guide.
Replace filter.
Clean with isopropyl alcohol
(or equivalent) or replace
light guide filter if it can not
be cleaned.
Properly install
lamp/reflector assembly.
13