Testing the BlueWave
1. Attach the detector to the radiometer (Figure 15).
2. Attach the BlueWave® QX4™ Optic Adapter to the detector (Figure 16) using the two 2-mm screws
provided.
Note: The Optical Adapter and Optometer are calibrated as a set, so they are not interchangeable among
other similar type radiometers.
Figure 15. Attach Detector
3. Insert the LED Wand into the Optic Adapter until it bottoms out. Tighten the set screw on the top of the
Optic Adapter when the LED Wand is installed.
Note: Do not over-tighten the set screw, as damage to the LED wand may occur.
Figure 17. Insert LED Wand into Optical Adapter
4. Press and release the On/Off button on the radiometer's face plate to turn the unit on.
5. On the radiometer's display, the top line displays the light source and the second line displays the mode.
Dymax ACCU-CAL™ 50-LED Radiometer User Guide
QX4
®
Figure 16. Attach Optic Adapter to Detector