Chapter 5
Control Solution Test
Control solution is used to check the performance of the kit. The meter performs adequately
when control test falls within the range listed on test strip vial label.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
control range level 0/ level 1/ level 2
G1303190
Test Strip
2017/01
98-164
mg/dL
5.4-9.1
mmol/L
Storage &
Storage &
Operation
Transportation
Transportation
90%
10%
EA13120901
2017-01
Storage &
Storage &
Operation
Transportation
Transportation
90%
10
10%
19
98-164
mg/dL
5.4-9.1
mmol/L
EA13120901
2017-01
Directions:
• Insert an unused strip
1
 into the meter.
• Shake well and discard
3
 the first three drops of
 control solution. And
 put one drop onto a
 clean surface.
• The result will display
5
 after countdown.
• Compare the result with
 the range listed on test
 strip vial.
New Taipei City 242
MDSS GmbH
Schiffgraben 41
30175 Hannover, G
FIG-80-SLG10-6BK
Control Range
level 0 / level 1 / level
98- 164
mg/dL
5.4-9.1
mmol/L
Storage &
Storage &
Operation
Transportation
Transportation
90%
10%
2
• Press "M" button for 3 seconds
 to enter the control solution
 mode.
4
• Touch the control solution with
 strip sample inlet .
Note:
• Do not reuse the test strip.
• Newly opened vial of control solution and test
 strips must be marked with the opening date.
• Close the strip vial lid completely after use.
• Do not use the meter if the control range do
 not fall in the correct range.
• If the control test keeps result wrongly,
 please contact an authorized distributor for
 help and service.
• Control solutions are not included. Please
 contact an authorized distributor for checking
 the meter kit performance.
20