- ページ 8

医療機器 Active Motif 53082のPDF マニュアルをオンラインで閲覧またはダウンロードできます。Active Motif 53082 15 ページ。 Cooled sonication platform

B. Placement of the Converter/Probe
To ensure sample-to-sample consistency, the platform
has a bracket secured near the top of the rod that holds
the sonicator converter/probe. This bracket enables
you to fix the placement of the probe tip. Depending
on the probe to be used, you may need to adjust where
the converter bracket is located vertically on the rod.
Typically, the black Stop Collar is placed on the top gear
of the rod, and the bracket is secured just above the
collar (right). (The Stop Collar is used to set how far up
the rod that the platform can be raised. Remove the pin
so that you can slide the collar to its desired height, then
reinsert the pin to fix the collar at the chosen rod gear.)
Once you have determined where the bracket will be
placed, use the supplied hex wrench to loosen the set
screws. Adjust the bracket to its new height and use the
hex wrench to tighten the set screws so that the bracket
will not move on the rod.
You will also need to determine at what height in the
bracket the converter/probe will sit. It is suggested that
you place tape or some kind of marking on the converter
(right). This will enable you to return it to the exact same
position in the bracket in the event somebody moves
(or removes) the converter/probe from the bracket.
Once the converter/probe is placed into the bracket,
secure by tightening the screw to the converter size
adapter. DO NOT OVERTIGHTEN. The bracket should
firmly secure the converter/probe to the platform, but
care must be taken not to damage the converter.
Note:
Depending on the size of your
converter/probe, you may want to
remove the converter/probe bracket
and replace it with the alternate spring
bracket. (See Appendix Section A)
www.activemotif.com
5