BASIC SETUP
Quick Setup Procedure
The steps below show how to set up and perform an operation check on
the Model 4500.
1. Inspect the 4500 and note the 5D module location placement at the
rear panel of the unit. Slot 1 is located at the far left moving towards
the right to Slot 4.
2. Wire the sensors to the appropriate 5D conditioning modules. Follow
the wiring configuration instructions laid out in the Model 4500 User
Guide that comes with the purchase of this instrument. Refer also to
the appropriate 5D manual for more information; the 5D manuals are
available from Daytronic or visit the website at www.daytronic.com.
NOTE: The sensors do not need to be connected in order to perform
this quick setup and operation check procedure; however, it is
recommended that a 5D conditioning module be installed to the unit
in order to stimulate input signals and verify that the proper sensors
respond to the appropriate data channels being configured.
3. Connect the DC power to the unit at 24Vdc at 1.5 amps nominal.
Ensure that the shield wire is tied to ground.
4. On power up, the Home page should be displayed and an audible
"beep" should be heard. The "beep" indicates that the A/D
11 |