- ページ 11

コントロールユニット DayTronic IL70AのPDF マニュアルをオンラインで閲覧またはダウンロードできます。DayTronic IL70A 15 ページ。 In-line dc strain gage conditioner

3. CALIBRATION
If the transducer calibration is unknown, the following calculation is useful to calculate an Equivalent Input
value for the factory installed, 59K Ohm shunt resistor. This will approximate the value assuming that the
mv/V sensititivity and bridge resistance of the sensor are known.
Sample Calculation: Assume that K = 3.000 mV/V for a 5000-pound loadcell
(fullscale) with a bridge resistance of 350 ohms.
Remote Calibration Check: The instrument can be placed in the calibration mode
(positive or negative) by shorting pin Power Common and SHP for positive or SHN for negative on the
lower I/O connector. Figure 4 indicates two methods of remotely entering the calibration mode
(external switch, or push button control). The Remote Cal function provides a convenient
method of periodically monitoring calibration of the instrument in the positive or negative realm.
DC Strain Gage Conditoner Module
(cont.)
25000 R
X =
K R
X =
where
Equivalent Input, % of full scale
R
=
bridge resistance, ohms
b
K =
transducer sensitivity, mV/V full scale
R
=
calibration resistance, ohms (59 k installed)
c
25000 x 350
X =
59000 x 3
Switch 1 - Mode -
Switch 2 - Volts -
Switch 3 - Current -
Switch 4 - Filter -
Switch 5 - Filter -
Switch 6 - Excitation -
Switch 7 - Zero Adj. -
Coarse Zero -
control with default calibration indicator.
Fine Zero -
Coarse Span -
with default calibration indicator
Fine Span -
Symmetry -
Over Range -
indicates the power input voltage is ON
Power -
Fig. 2 Front Panel Settings and Indicators
b
c
49.4%
=
of full scale = 2472 pounds
current (I) or voltage analog output
+/- 5 or +/- 10 Vdc when mode is voltage
4-12-20 or 4-20mA when mode is current
5 kHz or 200 Hz at 3 db, for 200 Hz switch 5 must be down
10 Hz or 200 Hz at 3 db, for 200 Hz switch 4 must be down
excitaton level, 2.5 or 5.0 Vdc - effects mV/V gain
extended (100%) or normal (20%) zero authority
16 position switch adjustment for stepped zero balance
16 turn potentiometer for fine zero balance control
16 position switch adjustment for stepped gain contr ol
16 turn potentiometer for fine gain - span control
adjust the negative output span to be equal to the positive output
indicates the analog output is 2% greater than mode selected
IL70A