AC配電 DayTronic IL78AのPDF マニュアルをオンラインで閲覧またはダウンロードできます。DayTronic IL78A 17 ページ。 In-line ac strain gage conditioner ac strain gage conditioner module
Shunt Resistor Selection
The IL78A has a Internal and External shunt
selection jumper located between the two wiring
connectors. Below are the appropriate wiring and
jumper selections depending on the application
using the recommended 7 wire configuration.
When more than one IL78A is being used in the same measurment setup, beat frequencies can be produced by
the excitation clock oscillator. To prevent this, the user should declare one of the IL78A units as a Master via
switch 6 on the front panel. The other units should be switched to Slave and wired as shown with the SYC
terminals connected and a separate Power Common wire. If the units share a common power supply, the
Power com terminal should be connected at the modules.
AC Strain Gage Conditioner Module
Fig. 2 Master / Slave Wiring
TRANSDUCER CONNECTIONS
IL78A
IL78A.2