- ページ 5

データロガー Dynamax SAPIP-LTEのPDF クイック・スタート・マニュアルをオンラインで閲覧またはダウンロードできます。Dynamax SAPIP-LTE 10 ページ。

Refer to USB Command List downloaded earlier.
Change logging interval to Test Normal (Type "INTER TN"), "TEST" and "Test Fast TF" mode are not avail-
able for LTE units
Change heater voltage to minimum required voltage for the sensor used. For example, if dynagage, set
heater to 3.5V (Type "AVRS 3500").
Connect sensor to SapIP then reboot SapIP. Wait for the reboot process to complete (about 60-90 sec-
onds).
Enter USB command mode then start logging (Type "START").
Monitor transmission on TeraTerm while verifying data on agrisensors.net (Test Normal logging interval
produces a new data every 100 seconds, it takes about 15 minutes for new data to be uploaded to agris-
ensors.net website).
While waiting for the new data, verify remote commands on agrisensors.net SapIP control panel page.
Monitor SapIP response through TeraTerm.
Status
Repeat Last Reading
Show Device Config
Update Date/Time
Send SMS to Device (or send a single letter from cellphone to the listed SapIP number)
SapIP-LTE doesn't wake up on demand. By sending 2 consecutive remote commands, SapIP-LTE will en-
ter debug mode and stay awake for 20 minutes. During "No Sleep" period, SapIP-LTE will response to
any remote command immediately.
When new data is ready and uploaded to website, verify it on agrisensors.net under "Sensor Report".
Stop logging by issuing remote "STOP LOGGER" command or typing "STOP" in TeraTerm.
SapIP-LTE is now ready to be deployed to the field.
Section 4 — Verify Log and LTE Communication