ラボラトリー機器 89 North PhotoFluor IIのPDF 交換マニュアルをオンラインで閲覧またはダウンロードできます。89 North PhotoFluor II 8 ページ。 Lamp replacement

89 North PhotoFluor II 交換マニュアル
STEP 4
Tabs (2x)
Retaining
wires (2x)
Lamp power cable in
upright (12 o'clock)
position
Lamp power connector
STEP 5
STEP 6
Rev B
INSTALL NEW BULB
While holding the white ceramic lamp base (Note:
do not touch reflector surfaces), place the new
lamp flush against the lamp mount's recessed
circular opening. The lamp power cable that
connects to the rim of the lamp must be placed in
the upright (12 o'clock) position. See Figure 3.
Try to move the lamp side to side to assure it is
properly seated in the recessed circular opening.
Figure 3: Top view of lamp chamber
ENGAGE the metal retaining wire and place them
within the tabs. The "spring action" wire will secure
the lamp in place.
CONNECT the lamps power cable (brown
connector). See Figures 3 & 4.
6