OPERATING INSTRUCTIONS
1. Connect either end of the 6" silicone tubing to the top of
Suction Canister connector, then connect the other end
to the bacteria filter. Use the other 6" silicone tube to
connect the bacteria filter to the Suction Unit (Figure 6).
2. The patient tubing connects to Suction Canister (Figure 1).
3. Make sure the power switch located on the side of the
unit is in the "Off" position before connecting the unit to a
power source (Figure 2).
Fuse
4. Turn the power switch to the "I" position, to start the device
(Figure 3).
5. Adjust the suction level from 0 to 560 mmHg by turning
the vacuum adjustment knob clockwise to increase
and counterclockwise to decrease (Figure 4). It is
recommended to test the pressure before using the
device on a patient. Simply block the end of the patient
hose or block the Suction Canister lid intake and allow
the gauge to reach a stable vacuum reading.
Vacuum Gauge
Vacuum Adjust Knob
2
Power switch
"Off" position
Figure 2
Power switch
"On" position
Figure 3
Figure 4
6. It is recommended to test the connections for any leaks
prior to being used on a patient.
7. Place the tubing to the appropriate location and start to
gently suction.
8. When the Suction Canister reaches a safe, full level
(800 cc), the suctioning will stop automatically. The
float shut-off, located under the lid, will activate to avoid
over-filling (Figure 5).
Suction Canister
9. Turn the power switch to "O" position to turn off the
suction motor.
10. Remove the lid and dispose the waste liquid from
Suction Canister.
CAUTION:
1. When the automatic float shut-off is activated,
contents in the Suction Canisters should be
disposed of. Further suctioning could cause
damage to the vacuum pump.
2. Should fluid be aspirated back into the unit, call an
authorized service technician to inspect the vacuum
pump as it may have been damaged.
Suction Unit • Vacuum Levels up to 560 mmHg (Reorder No. 32300-32301)
Figure 5
Safe
Full
Level