and/or patient. 
3. Squeeze the resuscitator bag several times to assure that the unit is functioning properly. 
4. Select the Peep Valve with appropriate range (2 ‐10cm H
5. Set the Peep Valve knob to the approximate value shown on the Peep Valve housing. 
6. Connect the Peep Valve to the Flow Diverter port as shown in the illustration. Connect 
the resuscitator to a manometer and test lung. Ventilate the test lung and adjust the Peep 
Valve as required to obtain the desired end expiratory pressure. 
7. Ventilate the patient following ACLS procedures for ventilation. Watch the chest rise 
during ventilation. Also, during ventilation, periodically check for: 
‐ signs of cyanosis; 
‐ adequacy of ventilation; 
‐ proper airway pressure; 
‐ secure connection of Peep valve and diverter. 
8. Between patient use or periodically, clean and sterilize the Peep Valve and Flow Diverter. 
‐ Testing the Resuscitator 
 
 
 
3.2 System Overview   
 
3.2.1 Intended for use together with other devices: 
Oxygen Tubing, Airway, Silicone Mask, Aircushion Mask, PEEP Valve, Disposable manometer 
 
3.3 Specification/Functions 
 
3.3.1 Operating Environmental Temperature Limits: 
‐18°C to +50°C 
3.3.2 Storage Environmental Temperature Limits: 
‐40°C to 60°C (‐104°F to 140°F) 
3.3.3 Range of setting: 
 
 
Valve resistance 
Oxygen 
concentration 
Bag capacity (ml) 
Max delivered 
volume (ml) 
Safety pressure‐
relief valve 
Reservoir bag (ml) 
Mask size 
 
 
 
Durable PEEP Valve 
REF 
SAP REF 
2403 
AV0035 
 
Adult Resuscitator 
1500 
1350 
60±10 cmH
2
2000 
#4 
Product name 
Durable PEEP Valve 
O or 5‐20cm H
2
Child Resuscitator  Infant Resuscitator 
<4.0 cmH
2
>95% @15 LPM 
550 
350 
40±5 cmH
2
1000 
#3 
Pressure range    Connector size 
2‐10cmH2O 
30F (Orange) 
O) 
2
280 
160 
40±5 cmH
2
500 
#0 
Package 
1 pc/pkg, 50 
pcs/box