Principle Drawing (B) – Inspiration
When compressing down the Resuscitator, it create the positive pressure and close the Intake Valve (G), the
air inside the bag pushes the Duckbill Valve (E) downward, and block the expiration port (F), and deliver the
air into the Silicone Bag then to the patient through the center of the Duckbill Valve, if the Oxygen is in use,
it should be connected by (H) part, then the Oxygen will fill up the Reservoir through the Reservoir Valve,
and installs in the Silicone Bag through the recovery inhale motion, then send directly into the patient's
body by compressing the silicone bag.
(E)
Principle Drawing (C) – Exhalation
When releasing the Silicone Bag (B), push upward the Duckbill Valve and keep it in close position, so to
release the exhale air through the Exhale Valve (F).