Basically, the unit consists of two major components, the Electronic
Generator and the Handpiece Assembly (Water Source is also NECESSARY
but not supplied with this model). The Water Supply is used from the
Dental operatory, and must be applied/configured by a qualified technician.
ELECTRONIC GENERATOR AND WATER SYSTEM
The electronic generator produces the frequency and the power to activate a
wide variety of interchangeable INSERT TIPS. A Water System/SUPPLY MUST BE
ATTACHED to your unit to provide precise control over the water flow which
serves to wash debris from the work area, and to cool the laminations of the
insert tip. The following is a brief description of controls needed for the proper
AUTOSCALER®:
operation of your
The AIR SOURCE (20 PSIG) minimum is required, and (must be
installed by a qualified technician) providing an independent supply of
air to your AUTOSCALER® unit.
POWER CONTROL KNOB: The linear Power Control knob adjusts the
output power to the insert tip (must be installed by a qualified technician).
WATER CONTROL (supplied by the dental operatory, and is not part
of the AUTOSCALER® unit) is for adjusting the flow of water to the
handpiece, and insert tip. To obtain the NECESSARY water flow of at
least (18cc - 30cc/+/minute) - i.e. a very strong mist, it may require more
than ONE or MORE complete rotations of the knob in either direction.
Depending on the water regulator supplied in your operatory the following
applies: HIGHER water flow to the RIGHT, and LOWER water flow to the
LEFT. Some water regulators adjust the opposite (more flow to left and less to
the right). Become familiar with your controls first before performing any
dental procedures on patients.
HANDPIECE ASSEMBLY
The handpiece assembly is used to accept Cavitron® interchangeable
insert tips. It is connected to the AUTOSCALER® through specially
routed wires in your operatory (INSTALLED BY QUALIFIED TECHNICIAN
ONLY). The open end of the handpiece will accept all Cavitron® brand:
"P", "TFI™", "SLIMLINE™" or "FSI™" style 25,000Hz (25K) inserts. If you
ordered a 30,000Hz (30K) model AUTOSCALER® you must use the
shorter (30K) frequency insert tips.
30K tips measure (Shorter style approx. 6.25 inches long)
25K tips measure (Longer style approx. 7.25 inches long)
*Cavitron®, SLIMLINE™, FSI™, and TFI™ are registered trade marks of Dentsply Sirona
30K
25K
19
(25K) inserts will NOT work in your 30K unit.
(30K) inserts will NOT work in your 25K unit.
Inserts are extremely FRAGILE . If one is dropped, bent, or scratched it
WILL NOT cavitate (i.e. will not function) or function by operating at a
higher than normal temperature and should be replaced immediately.
Insert tips should be routinely replaced according to insert
manufacturer's recommendations.
ELECTRICAL REQUIREMENTS
A GROUNDED electrical receptacle is required supplying 120-130
volts 50/60 Hz A.C. If a 3 to 2 prong adapter is used, the ground wire
MUST be connected to the base plate screw. Have an electrician test
your outlet for correct wiring polarity and a safe ground. DO NOT plug
the AUTOSCALER® in a GFI (Ground Fault Interrupt ) outlet.
HANDPIECE—POWER CONTROL POT INSTALLATION
Two 18 gauge wires must be routed between the AUTOSCALER®
console and the handpiece assembly. Thinner wire may result in
poor tuning/operation or non function.
Two 18 gauge wires must be routed between the AUTOSCALER®
unit and the Power Control Pot. A minimum of ½ watt rating @ 5K
ohms is required if substituting the power control pot with
another brand/style.
INSTALLATION AND OPERATION
1. The Water Control Knob should be in the HIGH position. It is
normal for the Knob to turn past the High position several
rotations which will increase water flow substantially.
Depending on the water regulator supplied in your operatory the
following applies: HIGHER water flow to the RIGHT and LOWER
water flow to the LEFT. Some water regulators adjust the opposite
(more flow to left and less to the right). Become familiar with your
controls first before performing any dental procedures on
patients.
2. Your unit is equipped with a AIR SWITCH ACTIVATOR. The air
supply must be attached to the rear of the unit, and a minimum
of ( 20 PSIG ) is necessary to activate the switch. DO NOT
REMOVE THE INTERNAL AIR RESTRICTOR FITTING. The restrictor
relieves the air input line if accidentally charged from another air
20