- ページ 6

アクセサリー Galltec+mela DWのPDF マニュアルをオンラインで閲覧またはダウンロードできます。Galltec+mela DW 10 ページ。 Sensors for humidity and temperature measurement with calibrated dmodul

Galltec+mela DW マニュアル
Connection diagrams
DWF 4...20 mA
DKF 4...20 mA
1
2
output humidity
3
+ UB
DKK 2x 0...10 V
DWF 0...1 V DC
DIF 0...1 V DC
1
GND
1
GND
2
+ U
15...30 V DC oder 13...26 V AC
B
2
+ U
6...30 V DC or 6...26 V AC
3
Ausgang Feuchte
B
3
output humidity
4
Ausgang Temperatur 0...10 V
4
5
GND
6
GND
DWF 0...10 V DC
DIF 0...10 V DC
1
GND
2
+ U
15...30 V DC or 13...26 V AC
B
3
output humidity
4
ESD protection advice
The sensors of the D Series contain components, which can be damaged by the effects of electrical fields or by charge equa-
lisation when touched.
The following protective measures must be taken when the housing of the sensor is to be opened for connection or in situ
alignment:
Before opening the housing of the sensor, ensure electrical potential equalisation between you and your environment.
Pay particular attention to ensure that this potential equalisation is maintained while you are working with the
opened housing.
Galltec+Mela
D Series_DW
page 6 of 10
4...20 mA
10...30 V DC
ensure galvanic isolation
from the power supply
0...10 V
0...1 V
0...10 V
DWK 2 x 4...20 mA
DKK 2x 4...20 mA
1
output temperature 4...20mA
2
output humidity
3
+ UB
DKK 2x 0...10 V
DWK 2 x 0...1 V DC
DKK 2x 0...1 V
1
GND
2
+ U
15...30 V DC oder 13...26 V AC
1
GND
B
3
Ausgang Feuchte
2
+ U
6...30 V CC ou 6...26 V CA
B
4
Ausgang Temperatur 0...10 V
3
sortie humidité
5
GND
4
sortie température 0...1 V
6
GND
DWK 2 x 0...10 V DC
DIK 2x 0...10 V
1
GND
2
+ U
15...30 V DC or 13...26 V AC
B
3
output humidity
4
output temperature 0...10 V
4...20 mA
10...30 V DC
ensure galvanic isolation
from the power supply
0...10 V
0...1 V
0...10 V
www.galltec-mela.de