EXCHANGING
BATTERIES
Batteries are inserted into the compartment in
the bottom side of the instrument.
To change the battery:
•
Half turn the screws on the compartment
cover and then carfully life the cover.
•
Make sure that the rubber sealing of the
cover will not be damaged or removed from
the guiding notch.
•
Insert the new battery.
WARNING! Ensure correct orientation
according to the sketch on the battery cover
to avoid internal damage to the tester.
When exchanging batteries, the memorized
parameter and calibration values are deleted.
Therefore you need to execute a new
calibration and enter the relevant parameters
again after exchange of the batteries.
Focus Ring
135˚ Swivel Lens
RUGGED
AUTHORITY
After battery replacement, switch the
instrument on/off twice to reset it to a
predefined operating status. All previous
settings will be restored to the tester after
connecting to the CorDEX CONNECT™
software, or the UT5000 Mini app.
ATTENTION! Only use a 3.6V Lithium Thionyl
Chloride Cell. The use of different batteries is
strictly forbidden.
OPERATIONAL
CAUTIONS
The cautions contained in this section do not
relate specifically to safe use in hazardous
area applications but should be followed
to help prevent damage to the device or
personnel.
•
Keep out of reach of children
•
The UT5000 is a precision instruction. Do not
drop it, strike it or use excessive force when
handling
Maintenance
For safety critical maintenance, please refer
to EN60079:17
UT5000 – Intrinsically Safe Ultrasonic Thickness Gauge
5