- ページ 11

計測機器 Decagon Devices AquaLab 4TEのPDF 据付時適格性確認/運転時適格性確認手順書および指示書をオンラインで閲覧またはダウンロードできます。Decagon Devices AquaLab 4TE 15 ページ。 Aqualab series 4, benchtop water activity meter

Standard @ 25°C
13.41m LiCl 0.250
0.250
8.57m LiCl 0.500
0.500
6.0m NaCl 0.760
0.760
1.00
Steam Distilled H
O
2
1.00
Water Activity Verification Standards Within Specification
Remarks:
N/A
If verification standards are out of specification, clean the instrument and follow the procedure in the
User's Manual to perform a linear offset. Repeat the verification process with fresh standards.
Standard @ 25°C
13.41m LiCl 0.250
0.250
8.57m LiCl 0.500
0.500
6.0m NaCl 0.760
0.760
1.00
Steam Distilled H
0
2
1.00
Water Activity Verification Standards Within Specification
Remarks:
N/A
Qualified by:
Date:
Lot #
Lot #
Initials:
Decagon Devices, Inc
Dew Point Sensor
a
±0.003
°C
w
Dew Point Sensor
a
±0.003
°C
w
June 28, 2016 11:24 AM
Volatiles Sensor (TEV only)
a
±0.015
w
Yes
No
Volatiles Sensor (TEV only)
a
±0.015
w
Yes
No
°C
°C
N/A
11