7.13. RANGE
Press the RANGE key to activate the manual mode and to disable the
Autorange function. The symbol "AUTO" disappears from the upper left
part of the display. In manual mode, press the RANGE key to change
measuring range: the relevant decimal point will change its position. The
RANGE key is not active in positions
Autorange mode, the instrument selects the most appropriate ratio
for carrying out measurement. If a reading is higher than the maximum
measurable value, the indication "0L" appears on the display. Press and
hold the RANGE key for more than 1 second to exit the manual mode and
restore the Autorange mode.
7.14. PEAK/INRUSH
1. In AC voltage test mode, Press PEAK/INRUSH key, the peak maximum
and peak minimum values can be measured.
2. In current test mode, Press INRUSH key the inrush current values are
measured.
7.15. Relative/Backlight
1. Press the REL button to zero the display. "REL" will appear in the
display. The displayed reading is now the actual value less the stored
"zero" value.
2. Press the REL button to view the stored value. "REL" will flash in the
display.
3. To exit this mode, press and Hold the REL button until "REL" is no
longer in the display.
4. DCA measurements Press the REL button to zero the display
5. Press and hold the REL/Backlight button to turn the Backlight on.
Press and hold again to turn the Backlight off.
7.16. Automatic Power OFF
1. In order to conserve battery life, the meter will automatically turn off
after approximately 15 minutes. To turn the meter on again, turn the
function switch to the OFF position and then to the desired function
position.
2. Press and hold the MODE/VFD key and turn the selector switch to turn
the system on, the auto power off function will be cancelled.