Calibration is used to check the potentiostat's health, and to "zero" many of the measurement circuits to
your laboratory environment. A warning does not necessarily indicate a critical failure, and Gamry can
use calibration information to determine the source of the warning.
Double-check the following:
Confirm all connections for the 200 Ω Calibration Cell.
Floating-ground cable is connected to calibration shield.
If possible, the Chassis Ground is attached to a known, good earth ground.
Click the Retry button, and the rest of the calibration restarts. Click the Ignore button for any other
calibration warnings that may appear, and continue to Step 3.
After a failed calibration attempt:
Find Calibration Results PC6-#####.txt in your My Gamry Data folder.*
E-mail the file, along with complete contact information, to [email protected]
* ##### is the serial number of your potentiostat.