ラボラトリー機器 CORNING LSE 6783のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。CORNING LSE 6783 9 ページ。 Digital water bath
CORNING LSE 6783 にも: 取扱説明書 (8 ページ)

CORNING LSE 6783 取扱説明書
14.0

TEMPERATURE RE-ADJUSTMENT

Water bath has a software for temperature re-adjustment. You must first measure
temperature in the middle of the chamber or in the sample for a minimum of 2 hours, which
is required to stabilize the temperature. You must measure temperature with a certified
reference temperature measuring instrument with precision of 0.1°C or more. After 2 hours
or more read temperature on thermometer and compare it with temperature on LED
display. Difference between thermometer reading and display is the value, which you will
enter in the software.
Example 1:
Reference Temperature is 37.9°C, temperature on display is 37°C
Difference is 37.9 - 37 = 0.9
Value is 0.9 (value which you enter)
Example 2:
Reference temperature is 36.2°C, temperature on display is 37°C
Difference is 37 - 36.2 = -0.8
Value is -0.8 (value which you enter)
Procedure for Temperature Re-adjustment:
Press UP and DOWN key at the same time for more than 5 seconds – until Cor
(correction) appears on the display and DOT blinks.
With UP or DOWN key, define value (when DOT is blinking). Pay attention to the
sign of value (plus or minus). After 5 of defining the temperature – display will
automatically show temporary temperature.
15.0
PRODUCT DISPOSAL INFORMATION FOR CORNING CUSTOMERS AND
RECYCLERS
According to Directive 2012/19/EU of the European Parliament and
th
Council of 4
July 2012 on waste and electronic equipment (WEEE) as
amended, the Digital Water Bath is marked with the crossed-out wheeled
bin and must not be disposed of with domestic waste.
Consequently, the buyer shall follow the instructions for reuse and
recycling of waste electronic and electrical equipment (WEEE) provided
with the products and available at the following link:
www.corning.com/weee.
6