EPC Single Valve Controller
Alicat Scientific EPC-Series Pressure Controllers incorporate a digital pressure
gauge with a single control valve and circuitry. The integrated control
algorithm measures the pressure, compares it with the setpoint, and adjusts
the valve accordingly at 1000 times per second.
The response time of the system will depend on the operating conditions and
the PID tuning of the device.
EPC-Series Pressure Controllers are normally intended to control the process
pressure downstream of the controller. In order for this to occur, a supply
pressure greater than the full scale pressure control range of the device
should be connected to the "IN" port.
EPC-Series Pressure Controllers can also be configured as back pressure
controllers, opening to release pressure upstream of the controller, and
closing to increase restriction and increase pressure upstream of the
controller. EPC Back Pressure controllers will have a "DS" in the part number.
EPCD Dual Valve Controller
Alicat Scientific EPCD-Series Closed Volume Pressure Controllers incorporate
a digital pressure gauge with dual control valves and circuitry. The integrated
PID loop measures the pressure, compares it with the setpoint, and adjusts
either the Inlet or Exhaust valve accordingly at 1000 times per second.
They are designed with a feed port, a process port, and an exhaust port. This
allows the controllers to raise and lower the pressure of a closed system
within the operating range of the controller without wasting gas under
constant pressure conditions.
Connect your EPCD into your process via the "OUT" port of the
unit. This is the "Process" port.
Connect a supply pressure greater than the full scale pressure
control range of the device to the "IN" port.
The "VENT" port can vent to atmosphere or be tied to a vacuum
source as the application demands.
Note: The pressure at the vent port should be at atmospheric pressure or
below to allow the controller to be used over its full scale range.