- ページ 9

コントローラー Alicat Scientific EPC SeriesのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Alicat Scientific EPC Series 17 ページ。 Pressure vacuum controller

Alicat Scientific EPC Series 取扱説明書

Polling Mode

Controllers are shipped in polling mode with a unit ID of A, unless requested
otherwise. Each poll returns one line of data. To poll, simply enter its unit ID.
Poll the device:
You can change the unit ID of a polling device by typing:
Change the unit ID:
Warning:
!
device on a single COM port.

Streaming Mode

In streaming mode, your device continuously and automatically sends a line
of live data at regular intervals. Only one unit on a COM port may be in
streaming mode at a time.
Note: RS-485 does not support streaming mode
To put your controller into streaming mode, type:
Begin streaming:
This is equivalent to changing the unit ID to "
out of streaming mode, assign it a unit ID by typing:
Stop streaming:
When sending a command in streaming mode, the flow of data will not stop
while the user is typing. This may make the commands you type unreadable.
If the device does not receive a valid command, it will ignore it. If in doubt,
simply hit
and start again.
The default streaming interval is 50 ms, but this can be increased by changing
Register 91 while the device is in polling mode:
Set streaming interval:
[unit ID]
a
Example:
[current unit ID]@=[desired unit ID]
a@=b
Example:
Care should be taken not to assign the same unit ID to more than one
[unit ID]@=@
A@=@
Example:
@@=[desired unit ID]
Example:
@@=a
[unit ID]w91=[time in milliseconds]
Example: aw91=500(streams data every 500 ms)
(polls unit A)
(changes unit A to unit B)
(Begins streaming unit A)
@
(stops and assigns unit ID of A)
9
". To take the flow controller