- ページ 2
浄水器 Aqua Free Germlyser IL 1500のPDF 使用方法をオンラインで閲覧またはダウンロードできます。Aqua Free Germlyser IL 1500 2 ページ。
1.3 Scope of delivery
• Germlyser
®
IL 1500
• Transport box
• Cable tie for product return and box fastener
• Zip lock bag for used filter
• Product information, sticker for filter replacement
1.4 Reprocessing
IL 1500 is a high-quality, reusable
®
The Germlyser
customer at a fixed rental rate. Reprocessing is carried out exclusively by Aqua free.
2 Certificates / Declaration of conformity
The company Aqua free GmbH is certified according to DIN EN ISO 13485.
filters are
-labelled medical devices.
The Germlyser
®
Extract from the declaration of conformity
Aqua free GmbH, Winsberging 31, DE-22525 Hamburg:
The manufacturer declares for the product Germlyser
on medical devices and medical products of Class I the conformity with the basic requirements of Annex
I of the council directive 97/42 EEC of the European communities.
3 Labelling
Manufacturer label
D/GB
Product information material
Do not use if package
is damaged
4 Preparation of commissioning and use of product
Please disinfect your hands and wear disposable gloves during filter installation
Turn off general water supply before installation
4.1 Adapter and hose installation
For initial installation or maintenance – if appropriate connection options (hose / pipe ø 8 mm) are not
available
- Open the water line to the point where the filter is to be installed. Note: Residual water can escape.
- Install the suitable quick-release coupling on counterparts in the pipe system. Note: Pay attention the
correct seating of the seal.
- Cut back the connecting hoses to the appropriate length, keeping blue hose as short as possible
-labelled medical device that is made available to the
IL 1500 to Annex IX of the directive 93/42 EWG
®
SN
Individual product label
Medical device label
Non-sterile
NON
STERILE
- Install connection hoses as follows:
○
Push the red hose or existing ø8mm hose/pipe completely
into the quick release adapter (filter inlet side)
○
Push the blue hose completely into the quick release adapter
(filter outlet side)
○
Pull the pipes to check the connection
4.2 Filter installation
- Disinfect fittings with an alcoholic surface disinfectant
- Open the packaging and remove the filter
- Connect the filter inlet to the red hose or existing ø8mm hose/pipe
- Similarly, connect the blue hose into the filter outlet side
- Pull the pipes to check the connection
- Open water supply, check tightness of the connections
and water flow
4.3 Filter replacement documentation
- Note the date of next filter replacement on the sticker and
place it on the filter in an easily visible position –
but not on the barcode.
5 Removal of used filter / filter change
Please disinfect your hands and wear disposable gloves during filter installation
Turn off general water supply before installation
The water in the pipes must not be under pressure
- Disconnect the blue hose from the filter
(Caution: Residual water can escape).
The connection is released by pressing
the retaining element against
the connector.
- Place zip lock bag over the filter
- Disconnect the red filter inlet hose (as above)
(Caution: Residual water can escape)
- Hoses remain attached to the adapters
- Place the filter into the provided transport box
- Equip transport box with protective foil for safe transport
- Close transport box with cable ties
- Install a new filter (see point 4.2)