TrachFlush Quick Guide
Alarms and Information Signals
Alarms
Cuff pressure ≥ 70 cmH
Decrease cuff pressure
Cuff pressure > 50 cmH
Decrease cuff pressure
Cuff pressure < 5 cmH
Increase cuff pressure
Cuff regulation error
Inspect the cuff for damage
Power off not possible
Disconnect airway and cuff pressure
tube from device
Cuff deflated > 1 minute
Inflate cuff or disconnect and turn off
device
Power is disconnected
Re-connect the power supply
Service required
Disconnect device and send to service
Information signals
Cuff pressure tube disconnected
O
2
Connect Cuff Pressure Tube or
change the tube
O
No ventilator signal
2
Connect Airway Tube and ventilator
Return to Pre-Flush Settings
O
2
Make sure to return to Pre-Flush or
Pre-Spont Flush ventilator settings
Flush Control in progress
Wait until Flush Control is completed
Flush Control complete
Press
Reading ventilator signal
Device is reading ventilator signal for
changes
Cuff pressure incorrect
If clinically safe, adjust cuff pressure to
the range 15 – 35 cmH
Pressure above PEEP too low
If clinically safe, increase pressure above
PEEP to ≥ 10 cmH
to return to main screen
O
2
O
2
Page 4 of 4
Inspiratory time too short
If clinically safe, increase
inspiratory time to ≥ 1,33 s
Frequency ≥ 28 b/min
Frequency is too high for Spont Flush. If
clinically safe, reduce Rf and try again
Cuff pressure above target
Cuff pressure is above the target pressure,
please wait for TrachFlush to reduced
Unstable ventilator signal
Inspect ventilator and press
Flush button to try again
Settings OK
Please wait, Flush or Spont
Flush will start