- ページ 13

アクセサリー AW-Lake EDG2のPDF インストレーション・オペレーション&メンテナンス・マニュアルをオンラインで閲覧またはダウンロードできます。AW-Lake EDG2 18 ページ。 The edge flow sensor with bluetooth

AW-Lake EDG2 インストレーション・オペレーション&メンテナンス・マニュアル
Analog Output Type and Forced Outputs
(Figure 12)
The toolkit allows the user to select different Voltage output
ranges. This is done by selecting the pulldown menu and
choosing between the four options available. Click Set after
selection is made.
• 0-5V
• 1-5V
• 0-10V
• 2-10V
The toolkit also allows the user to force one or both analog outputs. This is done by checking
the box and entering in the desired value. Click Set when done to have outputs take effect.
*Note: information will not be stored until SET is pressed and forced value will revert back to previous
value or state if output value is outside of the range.*

Analog Output Calibration

Warning modifying these outputs could result in inaccurate analog outputs.
This menu is provided to adjust the EDGE analog output to correspond correctly with the
analog output of the input device.
* It is intended for use with slight output differences. These values
are factory calibrated before shipping to match meters that are
calibrated yearly. *
In the top left corner of the toolkit is the Setup tab. This is
where you will find the analog output Calibration. (Figure 13)
Channel A – Voltage Output
Channel B – mA Output
Figure 12: Forced output and Type
Figure 13: Output Calibration
13 |
P a g e