Problem
Bed does not operate
Bed operates on battery only > check that the bed is plugged in, and that the wall outlet is working.
Only a portion of the bed
operates, but others work
Scale does not operate
(OPTIONAL)
If none of these suggestions results in full function returning to the bed then please, refer to properly trained
service personnel for technical assistance.
WARNING: If the bed looses any function or is damaged in any way stop using the bed
immediately and refer it to a qualified service personel. Further equipment damage or personal
injury may result.
Check the following:
> check all cords for fraying, cuts or damage and replace immediately.
> check that the bed is plugged in, and that the wall outlet is working.
> check for green, power on light, on top of control box, this indicates AC power to
the box. If no light, and wall outlet is working, then fuses may be bad (requires tools).
> if all or some of the relays in control box "click" when pendant is activated, and
bed does not move, replace control box.
> if no relay "click" can be heard from control box, check that pendant and other
cords are completely plugged into control box.
> if control box is found to be bad, check that no cords have been pinched or
severed. Also check that actuators and all moving parts are not obstructed. Also
check for damaged or dented actuators.
> check all cords for fraying, cuts or damage and replace immediately.
> check that you are applying enough pressure to the pendant button to perform
the function selected.
> check that all plugs are securely installed into the control box and that the
control box plug clamp is in place.
> check that travel on all actuators and other moving parts are not obstructed.
> try different location on control box. (ie: if head gatch does not work but knee
gatch does, plug head gatch into knee gatch outlet. If head runs, then problem is
in control or pendant, if no function, bad actuator.
> if no display, check batteries for possible replacement need.
> check that the scale display is plugged into the load cells, and that the connector
to each load cell (at each caster) is properly connected.
> check that casters are properly aligned with bed and that caster brakes are set.
> check that the cabling on any of the interconnects is not cut or crimped causing
a signal problem.
> check that there is clearance between the top of the caster and the bottom of
the load cell housing. There must be some space between the top of the caster
bearing cover and the housing.
TRI-FLEX II BED - OPERATION MANUAL