6. After completing the inhalation period recommended by your doctor, switch the ON/OFF switch 3 to position «O» to turn off the
device and unplug it from the socket.
7. Empty the remaining medication from the nebuliser and clean the device as described in the section «Cleaning and Disinfecting».
CLEANING AND DISINFECTING
Thoroughly clean all components to remove medication residuals and possible impurities after each treatment. The compressor 1 and
the air hose 6 should be cleaned with a clean, moist cloth.
Always wash your hands well before cleaning and disinfecting the accessories.
Do not expose the compressor to water or heat.
Replace the air hose for each treatment with a new patient or in case of impurities.
Unplug the device before cleaning.
Cleaning with water
Wash all nebuliser components (except air hose) under warm tap water (max 60° C) for about 5 minutes adding if
necessary a small quantity of detergent following dosage and use limitations as provided by detergent manufacturer. Rinse thoroughly
making sure that all deposits are washed away and leave to dry.
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use limitations as
provided by disinfectant manufacturer. Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121° C (20 min.) or 134° C (7 min.). After
sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.
MAINTENANCE, CARE, AND SERVICE
Order all spare parts from your dealer or pharmacist.
It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization cycles. Check the filter
continually for cleanliness and replace it if dirty, or after a maximum of 3 months use. Spare filters are provided with the device. To
replace the filter, open the air filter compartment 4 marked «Filter» and change the filter 11.
MALFUNCTIONS AND ACTIONS TO TAKE
The device cannot be switched on
Ensure the power cord 2 is correctly plugged into the socket. Ensure the ON/OFF switch 3 is in the position «I».
The nebuliser functions poorly or not at all
Ensure the air hose 6 is correctly connected at both ends. Ensure the air hose is not squashed, bent, dirty or blocked. If necessary, replace
with a new one. Ensure the nebuliser 5 is correctly assembled 12. Ensure the required medication has been added.
TECHNICAL DATA
Piston electrocompressor with thermal cut
Nominal voltage: see data label
Dimensions: 310x130x220 mm
Weight: 2.0 Kg approx.
Temporary use: 30/30 min. ON/OFF
MDD 93/42/EEC - Class of risk IIa
Maximum pressure: 250 kPa
Air flow: 14 Lt/min
Noise level: 56 dBA (EN13544-1)
Neb rate: 0.45 ml/min (with 0.9% NaCl solution)
Nebulizer capacity: min. 2 ml - max. 12 ml
Operating pressure: 90 kPa (0.9bar)
Operating flow: 6 l/min
MMAD: 2.38 μm GSD: 2.84 Aerosol Output: 160 μl Output Rate: 100 μl/min
Operating Temperature/Humidity: +10° C to +40° C/10% to 95% RH
Storage and Transport Temperature/Humidity/Air Pressure: –25° C to +70° C /10% to 95% RH
7
/690 - 1060hPa