4
Connect cannula
to the oxygen port.
NOTE: the oxygen
port connects
magnetically,
this is for your
5
safety (figure A)
Place the nasal cannula in a
comfort able position, with the
prongs inside your nostrils
6
Press the blue Power button to
begin oxygen flow (figure B)
7
An extended beep
indicates your device
is powering on. Please
wait until start up
is complete before
changing any settings.
8
Your Aer X will now be
providing oxygen when
you inhale
QUICK TROUBLE
SHOOTING TIPS
If the 3B
AER X should shut down during
™
start up or, the Yellow light (Fig 6 ) is
illuminated after start up;
1. Check to make sure the 3B AER X cannula is
attached. USING A DIFFERENT CANNULA THAN
THE CANNULA SHIPPED WITH YOUR 3B AER or
HAVING NO CANNULA ATTACHED may cause the
3B AER X to shut down during startup.
2. Make sure that the battery is charged. A
DEPLETED BATTERY will cause the 3B AER X to
shut down during startup.
3. Power down your 3B AER X and repeat the
startup process.
If the 3B AER X fails to start again;
4. Disconnect the battery and connect the power
supply.
5. Repeat the startup process.
If the 3B
AER X continues to fail to start up, please
™
contact your equipment provider for assistance.