アクセサリー Burkert MS05のPDF マニュアルをオンラインで閲覧またはダウンロードできます。Burkert MS05 4 ページ。 Turbidity sensor cube

Burkert MS05 マニュアル
MS05
Type MS05
can be combined with...
Type 8905
Type 8920
Online Analysis
Communicator
System
This sensor cube measures turbidity ac-
cording to DIN EN ISO 7027 or EPA method
180.1 and is designed for operation on a
fluidic backplane in the Online Analysis Sys-
tem 8905.
The continuous analysis of turbidity in water
is an indicator of undesirable, undissolved
substances in water. The measurement
before and after filter stages can indicate
the filter effect and enables, for example,
the optimisation of backwashing processes.
In the best case, this can lead to water and
energy savings.
The electrical and fluidic connections are
made via the connection panel of the sys-
tem. The sensor cube communicates with
the system via büS, allowing fully automatic
login to the online analysis system. If the
sensor is plugged into the system, it is
included in the list of büS members and fur-
ther adaptations to customer requirements
can be made.
Turbidity Sensor Cube
• Fully compatible with büS systems and a wide range of further
analysis sensor cubes
• Optical sensor according to DIN EN ISO 7027 or EPA method
180.1
• Modular sensor cube for hot swap (exchange during operation)
• Minimal sample water flow needed
General data
Compatibility
with Online Analysis System Type 8905
(see corresponding data sheet)
Materials
DIN EN ISO 7027
with sensor acc. to
Housing / Lever / Seal
PPE+PS / PC / EPDM
Cuvette / Valve
Glass / Silicone
Electrical connection
Spring contacts in the fluidic backplane of the Type 8905
Fluidic connection
Via pinch valve in the fluidic backplane of the Type 8905
Turbidity sensor acc. to
Light scattering, replaceable cuvette
DIN EN ISO 7027
IR-Laser
EPA method 180.1
Tungsten lamp
Turbidity measurement
DIN EN ISO 7027
with sensor acc. to
Measuring range
0...40 FNU
Resolution
±0.0006 FNU
±0.02 FNU or 2 % of M.V.*
Measurement deviation
3)
(the greater applies)
Linearity
±0.5 % of full scale
±0.02 NTU or 2 % of M.V.*
Repeatability
(the greater applies)
Response time
Depending on filter settings
(t90)
(by default 8 samples = 1 s)
Maintenance
12 months nominal, depending on the water quality
Regular manual or automatic cleaning
Type of medium
Water without particles: drinking water, industrial water
pH value
pH 4...pH 9
Sample water temperature
+3...+40 °C
Sample water pressure
PN3
Sample water flow range
> 6 l/h
Sample water filter
> 100 µm
1)
Only for sensor acc. to DIN EN ISO 7027 and only by Bürkert qualified staff - contact your nearest Bürkert facility
2)
Further measuring ranges on request
="measurement bias" as defined in the standard JCGM 200:2012
3)
* M.V.= measured value
EPA method 180.1
PPE+PS / EPDM
PET, glass / Silicone
,
1)
EPA method 180.1
2)
0...40 NTU
2)
±0.005 NTU
±0.02 NTU or 2 % of M.V.*
(the greater applies)
±0.5 % of full scale
±0.02 NTU or 2 % of M.V.*
(the greater applies)
Depending on filter settings
(by default 8 samples = 1 s)
(with Type MZ 20)
(+37...+104 °F)
www.burkert.com
1/4