6.2.
installation of a 8200 g2" version
onto an s020 fitting
Make sure the S020 fitting used for the installation of a G2"
holder is a special fitting for analysis (see the data sheets of
the S020 fitting).
Fit the fitting at an angle of ±75° max. to the vertical in order
to ensure the good operation of the pH/redox probe.
→ Install the fitting in compliance with the instructions provided in
the Operating Instructions.
→ Check that the seal is on the fitting and
that it is not damaged.
Replace it if necessary.
→ Unscrew the nut 2 from the holder.
→ Insert the nut 2 on the fitting S020,
identified 5.
2
→ Insert the snap ring 3 in the groove 4.
3
→ Check that the seal 1 is correctly
1
4
inserted in the holder.
→ Place the holder onto the fitting.
5
→ Screw the nut 2 to the holder to fix it to
the fitting.
→ Insert the probes in the corresponding
holes.
Fig. 1 :
Installation of a G2" holder
10
English
Type 8200
Installation,commissioning
6.3.
installation of a 8200 g1" version
onto a Tee fitting
→ Check that the seal is on the holder.
→ Screw the G1" holder onto the Tee
fitting.
→ Insert the probes in the corresponding
G1"
holes.
Fig. 2 :
Installation of a G1" holder
6.4.
installation of a stick connection
8200 onto a Tee fitting
→ Stick the 8200 PVC-U holder onto
a PVC-U Tee fitting with appropriate
dimensions.
→ Insert the probes in the corresponding
Ø 32
holes.
Fig. 3 :
Installation of a stick holder