- ページ 4
データロガー Dickson SM320のPDF マニュアルをオンラインで閲覧またはダウンロードできます。Dickson SM320 4 ページ。 Temperature & temperature/ humidity data loggers
Dickson SM320 にも: はじめに (4 ページ)
1.800.561.8187
Troubleshooting
No communication - Serial Connection
• Verify that you have the correct version of
DicksonWare™ software installed on your PC. Version
8.0 or higher is required for models SM300, SM320,
SM325, TM320 and TM325.
• Verify that the correct COM port is selected: From
the main Dicksonware screen, click on Logger then
Communication, a black dot will appear next to the
selected COM port. You may need to select a different
COM port. Should you get an error message stating
that "Device Is Already Open", this could mean that
you have the proper COM port selected, but another
device, or it's software, has it allocated. Palm pilots,
for example, will cause this problem, in which case
the port is not actually "available" and you may have
to disable that device.
• You may need to relocate the download cable to
another serial port on the back of the PC and possibly
try changing the COM port again in DicksonWare™.
• If communication has not been established with the
previous steps, you may need to open the logger and
press the reset button and then try all COM port and
cable combinations again.
• If possible, try another PC
• Make sure the USB is unplugged, and that "USB" is
NOT checked in File > Preferences > Serial.
No Communication - USB Connection
• Make sure that "USB" is selected under File >
Preferences > Serial.
• Unplug USB cable and plug back in.
• Remove all power to the logger (this will not cause the
unit to lose any data within the logger, but
you will have to start the unit logging again using
DicksonWare™) unplug the USB cable, power the
logger back on, then reconnect the USB cable.
• If the logger was used in a moist or humid
environment Condensation may have formed on
Error Codes
Err 1 .........................No memory card
Err 2 .........................Memory card locked or protected
Factory Service and Returns
Contact the factory for a Return Authorization (RA)
Number before returning any instrument. The model
number, serial number and a purchase order number will
be requested before an RA number is issued.
CE testing has found that when using unshielded thermocouple probes in a strong RF field in the 350 to 400 MHz
frequency range, the reading may be off by several degrees.
www.
the unit. Place unit in a warm dry environment for
24 hrs. Clear memory and try again. These loggers are
designed for use in a non-condensing environment.
If the environment creates condensation, try placing
the unit (temperature only models) in a small sealed
plastic bag to protect it from condensation.
• If possible, try another PC, and/or another USB port.
• If the logger appears to be frozen or non operational,
try the following: Open the unit by removing the
screw on the back of the case and remove and
reinstall or replace the batteries.
NOTE: Removing the battery will not erase any stored
data
Unit will not turn on or frozen display
• Unit may be "locked up", try removing the battery
and then replacing after a few moments.
• Battery may be to low or dead, replace batteries
The display reads Prob on RF (remote probe)
• (This would be for SM320 and SM325's only), it
means that the unit does not recognize the external
probe, check to see that it is connected properly. If
it is connected properly, there may be a weak internal
connection and should be returned.
• Is there an extension cable being used? It could be
defective. Omit the extension cable and try another
K-type probe.
Channel shows extreme reading on the graph
• This is normal when an activated channel has no
probe connected at the time a sample is taken.
Err 23.......................Memory card requires reformatting
Err 66.......................Memory card full
NOTE: Dickson shall not be liable for consequential or
incidental damages resulting from failure or malfunction
of its products.
DICKSON
.com