Dickson ensures that the Replaceable Sensors / Instruments listed below were developed, tested
and validated together.
Replaceable Sensor Model: R200 and R250
Instrument Model: TH8P0, TH8P2, TH8P3, TH8P5
Operation
Operation
Operation
Operation: The Instruments and Replaceable Sensor models listed above were specifically
designed to work together. No other sensors will work with the Instrument models listed. No
other Instrument will work with the Replaceable Sensors listed. The Replaceable Sensor must be
connected to the Instrument for the Instrument to operate properly.
Calibration: Only the Replaceable Sensor is calibrated. The sensor, and all calibration defaults
Calibration
Calibration
Calibration
and adjustments are stored on the Replaceable Sensor. Accuracy is strictly controlled by the
sensor – no adjustments are made by the Instrument. Readings are sent from the Replaceable
Sensor to the Instrument for storage purposes only. Certificates of Calibration / NISTs are only
supplied for the Replaceable Sensor for this reason.
Recalibration
Recalibration: When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor
Recalibration
Recalibration
can be replaced with a new Replaceable Sensor thus eliminating the need to return the entire
instrument for recalibration. Replaceable Sensors can be returned for recalibration as well.
Dickson TH8P Manual V14 052418