Model DS1200-PDT - Dewpoint Hygrometer
Installing the Model PDT in a Air/Gas Sampling System - Continued
The piping installation schematic diagram below shows all components, which could be
used in a dry gas measurement application although not all the items shown will be
required for every installation.
The flow rate, although not critical to the sensor measurement, should be low enough
to avoid abrasion to the sensor surface without being so low as to extend the system
response time to an unacceptable level. In general, a flow rate of between 2 and 3
litres/minimum at NTP will give the right balance.
The sensor is a variable capacitor, which is directly affected by changes in partial pressure
of water vapour. These changes are proportional to the dew/frost point temperature.
The measuring transmitter can be installed directly into the process line or ducting, but
this can create problems with access for maintenance and calibration. It is for these
reasons that we recommend that the transmitter be installed in a bypass, fast loop or
total loss sample system where the transmitter is accessible without interrupting the
main process flow line.
Piping Installation Schematic